Product Name
|
Radial Artery Compression Devices
|
Appearance
|
Smooth appearance, soft, clear outline, no burr, flying edge, scratches, shrinkage, and other defects
|
|
No air leakage;
The connection between the fixed wrist strap and the fixed wristbands should be able to withstand 20N static tension for 15 seconds;
Capacity tolerance of radial arterial hemostatic injection cylinder: the maximum tolerance for more than or equal to half the nominal capacity is ±10% of the discharge volume;
The radial arterial hemostatic injection cylinder body should be well-sealed;
The horizontal adhesion force of the adhesive buckle of the radial artery hemostatic device should be greater than 20N;
The avulsion force required to open the buckle of the radial artery hemostatic device should be less than or equal to 15N.
|
Package
|
The package shall be in good condition and there shall be no channels or holes in the bonded sealing area
|
Sterility
|
Sterilized with ethylene oxide
|
Residual EO
|
After sterilization with ethylene oxide, the residual ethylene oxide should not be more than 10μg/g
|