Product Name
|
Single-Use Suture Grasper Device
|
Appearance
|
Smooth appearance, soft, clear outline, no burr, flying edge, scratches, shrinkage, and other defects
|
|
Press the button, sliding should be flexible, and clamp open and close freely;
Do the pull test under 10N tension, the needle and the handle should not be loosened or separated;
The needle shall have good corrosion resistance, and its corrosion resistance shall not be less than grade 5.4B of YY/T 0149-2006;
In the test according to Appendix C in GB/T 18457-2015, the difference in PH between the test solution and blank control solution should not exceed 1.0;
The test was carried out according to Appendix D in GB/T 18457-2015. The span was (30±0.1) mm, the load was (40±0.1) N, and the deflection value of the puncture needle should not be greater than 0.5mm;
The needle tip shall be sharp, without a bending hook, and the puncture force shall be less than or equal to 10N when tested
according to the method specified in Appendix A. 2.10 the packing seal. |
Package
|
The package shall be in good condition and there shall be no channels or holes in the bonded sealing area
|
Sterility
|
Sterilized with ethylene oxide
|
Residual EO
|
After sterilization with ethylene oxide, the residual ethylene oxide should not be more than 10μg/g
|