Customization: | Available |
---|---|
Type: | Basic Surgical Instruments Sets |
Material: | ABS+06cr19ni10 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Name
|
Single Use Suture Grasper Device
|
Appearance
|
Smooth appearance, soft, clear outline, no burr, flying edge, scratches, shrinkage and other defects
|
|
Press the button, sliding should be flexible, clamp open and close freely;
Do the pull test under 10N tension, the needle and the handle should not be loosened or separated;
The needle shall have good corrosion resistance, and its corrosion resistance shall not be less than grade 5.4B of YY/T 0149-2006;
In the test according to Appendix C in GB/T 18457-2015, the difference of PH between test solution and blank control solution should not exceed 1.0;
The test was carried out according to Appendix D in GB/T 18457-2015. The span was (30±0.1) mm, the load was (40±0.1) N, and the deflection value of the puncture needle should not be greater than 0.5mm;
The needle tip shall be sharp, without bending hook, and the puncture force shall be less than or equal to 10N when tested
according to the method specified in Appendix A. 2.10 the packing seal. |
Package
|
The package shall be in good condition and there shall be no channels or holes in the bonded sealing area
|
Sterility
|
Sterilized with ethylene oxide
|
Residual EO
|
After sterilization with ethylene oxide, the residual ethylene oxide should not be more than 10μg/g
|